Laser Alcon Pure Point

Endo Fotocoagulacion en retina

Ensure that all personnel are wearing protective eyewear, OD 4 or above

at 532nm.

For Slit Lamp: Adjust the Intra-pupillary distance and the biomicroscope

oculars focus so that the image is clear.

For LIO: Adjust the intra-pupillary distance on the headset so that the image

is clear.

Target the red aiming beam on the area to be treated.

Go to Ready mode by pressing the Standby/Ready Mode or footswitch

sideswitch (if enabled).

WARNING: Laser is ready to fire.

Press the footswitch when ready to fire.

Repeat the firing procedure as often as necessary, making adjustments to

power output and duration as appropriate to complete the treatment session.

When the treatment is complete, release the footswitch and press the

Standby/Ready key.

To shutdown system, turn key to OFF (O) position and rear panel

power switch to OFF (O)

Position.approximate Dimensions of the pure Point Laser console:

Width (overall)=0.23 m

Length(overall)=0.35 m

Maximum Height (overall)=0.18

Weight= <13.6 kilos

 

Characteristics:

Voltage:100-120 VAC 5A 220-240 VAC 2.5A

Frecuency: 50/60HZ

Ilumination power port output: 7.5V maximum at 2.25A